Your Location : Home > Products

ESR Fast Detector

ESR Fast Detector

Previous: Capillary

Next: ESR Tube Rack

Product Description

Main Features:
1. Better compatibility compared with Westergren method,
2. A safe, reliable and airtight operation during the whole process from blood collection to test, 
    free of biological contamination. 
3. Designed with spirít level of patent technology, which ensure the detector horizontal and aclinic.
4.10 channels, synchronous operation allowed,
5, Only 30 minutes needed for reading, easy and fast.
6. Stainless steel soleplate silicagel cushion on, which will avoid any noise or breakage when the ESR 
    tube being placed.
Howto Use;
1. Carry out the venous blood collection according to standard requirements with 8 x 120mm ESR 
    vacuum tube.
2 Immediately invert the tube at 180 degrees for 6 - 8times to achieve thorough mixing, which will 
   avoid hemolysis, dotting or blood bubble. 3, At the room temperature of around 20°C. vertically 
   place the ESR tube containing blood sample onto the detector, note down the starting time 
   and relevant numbers, Keep the detector still for 30 minutes and then read the millimeter of 
   erythrocyte sedimentation. 4, Detailed reading method; Keep it stable for 30min, align the 
   plasma concave in ESR tubes to the zero scare of the detector.
   Then, read the data by aligning the upper surface of erythrocyte to the scale on the 
   detector, (See as the schematic diagram)
Notice:
1. Before operation, the EST fast detector should be put on a steady platform and adjustable to 
    be horizontal with knobs below the soleplate. 
2. The ESR tubes should be kept vertical during the whole sedimentation process, angle of inlintion 
    should not be more than3 °C. Or else, the testing result will be of large deviation.
3. Keep the room temperature at 20=5 °C, sedimentation time 30t2min.

[ Return ]